The following table describes the meaning of each status for each indicator:
Table 1-1: Indicators
Indicator Status What it Normally Means
Standby/on indicator
O The device has no power.
Green The device is in standby and not emitting light.
Blue The device is emitting light.
Hour meter display
O The device has no power.
Displaying numbers
The device is in standby or emitting light. The
number represents the number of hours the device
has been emitting light during the the life of the
device.
Over/under temperature indicator
O No temperature problems detected.
Steady red Under-temperature condition
Flashing red* Over-temperature condition
*The over/under temperature red indicator briey ashes every time the Standby/On switch is depressed to turn
on the therapeutic light in order to conrm the function of the alarm indicator.
For operator troubleshooting instructions, refer to Chapter 4.