Chapter 2: Product Setup and Operation
Upon shipment, the light pipe is detached from the light box. Prior to rst use, the light pipe needs to be
attached and secured to the light box and then the device needs to be mounted on the dovetail rail of a
microenvironment or a portable roll stand.
Installing the device shall be performed by authorized service personnel only. For instructions, refer to
section 7.4.
WARNING:
To avoid patient injury, do not mount the system while a patient occupies the bed.
WARNING:
To avoid the risk of electrical shock, the system must only be connected to a supply mains with
protective earth, using the power cord supplied.
2.1 Pre-use Checkout Procedure
Follow below checkout instructions prior to rst system use and then before any time the system is put
into operation:
1. Examine the device for missing parts or obvious signs of damage, including power cord, light pipe,
light box, and labels. If damage is found, do not use the device and contact service personnel.
2. Check that the light box is securely attached to the accessory dovetail rail or roll stand, i.e. the device
does not slide up or down the dovetail rail.
3. Move the light pipe to verify that it moves freely and stays in position.
NOTE: The light pipe will typically experience some minor deection (less than 1.3cm) after positioning.
4. Connect the power cord to an appropriate power source. The standby/on indicator will light green to
indicate standby mode.
5. Press the standby/on switch to turn on the therapeutic light. The standby/on indicator will change
from green to blue and the over-temperature/under-temperature indicator will briey illuminate.
6. Conrm that the fan is operating by momentarily placing your hand close to the exhaust vent when
the device is on to feel the air ow coming out of the device.
7. Focus the therapeutic light such that the spot diameter is 35.5 cm (14 in) or the shade is 38 cm (15 in)
from the bed surface.
17