11.1 Air Filter Replacement
1. Press the left and right sides of the air lter guard inward to release the guard and remove the lter (Refer
to Figure 11-1).
2. To re-install, put the air lter on the vents. Place the guard on the lter and insert the tabs into their
corresponding slots on the system to hold the lter in place.
11.2 Fuse Replacement
1. Make sure the system power cord is unplugged from the power outlet.
2. Release the power cord retainer to unplug the power cord from the system.
3. Use a at-head screwdriver to open the fuse compartment lid of the power inlet module and pull the
compartment out to access the fuses (Refer to Figure 11-2).
4. Remove the fuses.
5. Replace these fuses with the two fuses (T3.15A @ 250V~, slo-blo type) in fuse kit 6600-0730-214.
6. Reverse steps to re-install.
CAUTION:
Always use the fuse type recommended by GE Healthcare for fuse replacement.
Figure 11-1
Figure 11-2
64
Chapter 11: Replacement Procedures