5 PIST0006 - Vascular Above Without Insert
5.1 Personnel Requirements
Personnel
Requirements
Preliminary Reqs Procedure Finalization
1 0 min 1 hour 0 min
5.2 Preliminary Requirements
5.2.1 Tools and Test Equipment
Item Qty Effectivity Part# Manufacturer
Carbide–tipped drill bit for
installing gantry and table
anchors: ANSI B94.12 or
5/8” for on–grade anchors
1 - - -
Carbide–tipped drill bit for
installing gantry and table
through bolts: 25 mm (1”)
recommended
1 - - -
10 mm (3/8”) drill bit for pilot
holes
1 - - -
Impact drill for use with
carbide–tipped drill bits. We
recommend Skill #906 roto–
hammer or equivalent.
1 - - -
1” wire brush for cleaning out
mounting holes
1 - - -
Vacuum cleaner 1 - - -
5.2.2 Consumables
Item Qty Effectivity Part# Manufacturer
grout + above grade floor
anchors without insert
1 - 2285051 -
Single Plane and Biplane Innova Systems Pre–Installation Kit Installation Procedures
GE Healthcare Direction 2290880-12-1EN, Revision 1