GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4) 8-7
8-3-4-1 Processes prior to R7.7.x Installation
8-3-4-2 Processes to install R7.7.x
Figure 8-113 Pre-installation Procedures
NOTICE
*1: Loading R7.7.x application software procedures requires, at minimum, C: Drives to be ghosted (as
opposed to full ghost). In such case, patient image data, preset parameters, and customer data will be
maintained. Nonetheless, for safe-keeping customer data, it is recommended to back up all data prior
to R7.7.x installation.
In case of Full Ghost, customer data/setting must be saved for restoration after the process.
Figure 8-114 Software Loading Procedures
NOTICE
*2: Do not power ON the peripherals when installing base system image. Disconnecting LAN is also
recommended procedure to ensure stable process.
Caution Item
Check SW
Level
POWER
ON
Save Customer
Setting
*1
POWER
OFF
Important Item
Power OFF
USB Devices
*2

Table of Contents