8-8 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-4-3 After installation of R7.7.x
Figure 8-115 Post Installation Procedures
NOTICE
*3: Verify that Probes are not connected to the scanner. Other peripheral devices such as printers
should be turned ON at this time.
*4: If Operator logon window appears, select "adm"(administrator) and enter the password that is
configured already (default value is no password ? just press "enter" key).
*5: If FE cannot see login window for Service Platform, then installation of the platform have failed. FE
must double check the installation setting, and repeat ghost/appli load procedure.
For entering password, make sure CAPS (lock) is OFF.
*6: Each computer name is unique. Check the scanner serial number printed on the label located at
the rear lower side of the scanner. For example, if 123456YM1 is printed, then enter LS6-123456YM1.
Regional Preset, Analog Peripheral, and UP-D897 setting must be required.
Make sure to press "Save Setting" before exiting scanner setup window, or otherwise operator cannot
enter the new patient screen.
*3
*4
*5
*6

Table of Contents

Other manuals for GE LOGIQ S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S6 and is the answer not in the manual?

GE LOGIQ S6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ S6
CategoryMedical Equipment
LanguageEnglish

Related product manuals