8-36 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-10 System Settings
After application is successfully installed, various LOGIQ™ S6 system parameters must be properly set.
Table 8-67 System Setting Summery
Item Method Proper Value
Software Version Utility
\
System
\
About
s/w version : R7.7.x
Image Part No. : 5305207
Computer Name Utility
\
Connectivity
\
TCP/IP
\
Computer Name
\
Enable DHCP
\
IP Address
\
Subnet Mask
\
Default Gateway
Same as Serial Number
OFF
restore customer setting
restore customer setting
restore customer setting
Patient
Database Check
Utility
\
Connectivity
\
DataFlow
\
Default Data Flow
Press Patient (Keyboard or Touch Panel)
Local Archive – Int. HD
(n/a – Automatic)
LCD Monitor Setting Press ‘Toggle Button’ for few seconds (various)
Software Option Utility
\
Admin
\
System Admin
\
HW Number Same as Option Dongle ID
VCR Utility
\
Peripherals
\
Video Setting
\
VCR
PAL/NTSC-depends on region
(set accordingly)
Language pre-work (For
Japanese Only)
Utility
\
Regional Option
\
Language (tab)
\
Install Files to East Asian Language
(Reboot)
ON / Apply