DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
8-64 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-11-1 Service Platform (cont’d)
4.) Select GE Service at the “Select User Level” field.
5.) Enter the password for the Service Platform.
6.) Click on Okay.
7.) Verify that the following screen (Service Platform) is displayed on the monitor.
8.) Verify that Service Platform Version is 2.2.0Q.
9.) Click on x located at the upper right corner of the service platform screen to close the Service
Platform and return to the scan panel.
CAUTION
If the Login window for Service Platform is not displayed on the monitor, the installation of the
Service Platform has failed. Reload the application software or Base System Software (OS) +
application software. Contact a Technical Support for details.
Figure 8-189 GE Service
NOTICE
When the service platform is NOT displayed, check if CAPS lock is selected. The CAPS should not be
selected.
Figure 8-190 Service Platform