DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
Section 10-7 - Electrical Safety Tests 10-29
10-7-9-5 Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current (cont’d)
8.) Test every probe and record in Table 10-26 the test results for every probe.
The test passes when all readings measure less than the values in Table 10-30.
9.) Keep a record of the results with other hard copies of Planned Maintenance data.
.
Table 10-26 Probes Tested for Isolation (Sink) Current