DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
4-16 Section 4-2 - General Procedure
4-2-6-3 Loading Presets from an MO/DVD-R disk
1.) Insert the MO/DVD-R disk with the archived Presets into the MO/DVD-R.
2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Restore
sheet will be shown on the monitor.
3.) Select the item to restore either from resource Files.
4.) Click on Restore. A message to make sure the restore process is displayed on the monitor. Click
OK. The restore status for each item is displayed on the Result column.
5.) Make sure “Finished OK” is displayed on the Result column.
Figure 4-27 Restore Sheet