GE COMPANY
D
IRECTION 5745295-1EN, REVISION 1 OPTIMA CT520 DOD IB UPGRADE INSTRUCTION
Page 32 Chapter 2 - Upgrade Procedure
Section 3.0
System Configuration
3.1 Restore System State
1.) Insert a previously saved system state DVD-RAM media into the peripheral tower drive or USB
media in any of the console’s USB ports.
2.) Wait until the media is ready.
3.) Select [Service] -> [Utilities].
4.) Select [System State] for DVD-RAM media or [System State - USB] for USB media. The
System State Save/Restore screen appears.
5.) Select [All].
6.) Select [Restore]. The Restore System State box appears.
7.) Select [Yes]. A message at the end of the System State Log window should display: Restore
System State Completed Successfully.
Note: If the site installed 6 options (refer to Section 1.1.5). Error message will be pop-up during this time,
please click [ok] to continue. Option Name is different between CTT system and SUSE system.
8.) When completed, select [Cancel].
Note: If Does Check option is installed, Dose Audit tool starts up when Reconfig Info or Protocol is
selected. That shows configuration and Protocol summary. Press [Quit] to continue.
9.) Select [Yes] when the Scan Hardware Reset pop-up appears.
10.) When completed, select [Dismiss].
11.) Remove System State media and place in a safe location.
12.) Verify Restore System State.
3.2 Configure System Information on Z820 (SUSE)
1.) Configure all remote network information.
a.) Select [ImageWorks] from the desktop.