IRECTION 5745295-1EN, REVISION 1 OPTIMA CT520 DOD IB UPGRADE INSTRUCTION
Chapter 2 - Upgrade Procedure Page 37
Section 6.0
Finalization
1.) Perform Quality Assurance Test of Functional Checks on Service Methods (5457941-2EN),
make sure all of them pass.
2.) Update GIB with the Product Locator Card information. Go to the following address to access
the site:
http://gib.gehealthcare.com/gib/gib_entry.jsp
3.) Insite Check out, refer to 5457941-2EN Service Methods CD.
4.) Save System State
a.) Insert the Save System State DVD into the Peripheral Tower DVD-RW drive.
b.) Select [SERVICE DESKTOP].
c.) If reloading software, select [UTILITIES].
d.) Select [SYSTEM STATE].
e.) Select [ALL] to select all the cals, characterization, etc.
f.) Select [SAVE].
g.) If applicable, select [OK] when the following message appears: System State Media
Status: Please insert a DVD into the drive and press Save again
h.) When completed select [DISMISS].
i.) Close the Service Desktop window in the upper left corner of the screen.