D
IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-75
PRELIMINARY
The selected data is copied. The backup process is indicated on the progress bar that appears.
6.) After the backup is complete, proceed to Software Update Procedure.
8-3-4-3 Software Update Procedure
1.) Turn OFF the system.
2.) Plug the Software Update Media into the USB port located on the Interface Panel of the Venue
Go™ .
3.) Turn on the system.
The opening screen is displayed.
4.) For Software Update: Select Update Venue Go SW to upgrade software without affecting user
data.
Figure 8-166 Copying Patients Dialog
Figure 8-167 Venue Go™ Software - Update Options

Table of Contents