IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-13
PRELIMINARY
8-2-6 WiFi module Replacement Procedure
8-2-6-1 Tools
Phillips screwdriver of appropriate size
8-2-6-2 Time Required
5 min
8-2-6-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown.
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
8-2-6-4 WiFi module Removal Procedure
1) Remove back cover.
2.) Remove the magnetic absorber (MAGRAM) cover from the top of the back-end, by gently lifting the
right side upwards, and gently pressing the connector knobs inwards (2 knobs in each side) -
especially the knobs on the left side (see below).
FRU Part #
Refer to Table 9-16 on page 9-17.
Back Cover Removal Procedure
Figure 8-85 MAGRAM cover