IRECTION 5813707-100, REVISION 2 VENUE GO™ SERVICE MANUAL
Chapter 3 - System Setup 3-15
PRELIMINARY
Section 3-5
Completing the Setup
3-5-1 Purpose of this Section
This section describes how to complete the setup of the Venue™.
3-5-2 System Specifications
3-5-2-1 System Requirements Verification
Verify that the site meets the requirements listed in Chapter 2.
(See: Facility Needs on page 2 - 7.)
Verify that the specifications below do not conflict with any on-site conditions.
3-5-2-2 Physical Dimensions
3-5-2-3 Mass with Monitor and Peripherals
3-5-2-4 Acoustic Noise Level
Less than 55 dB(A) at 20 degrees Celsius, measured in the operators head position, 20 cm in front of
the keyboard’s right corner, at 1.30 m above the floor, and in a distance of 1 meter at all four sides, 1
meter above the floor.
Table 3-13 Physical Dimensions of Venue Go™ System Only
Height Width Depth Unit
33 41 12 cm
13 16.14 4.72 Inches
Table 3-14 Physical Dimensions of Venue Go™ Mounted on Cart
Max. Height Max. Width Max. Depth Unit
125.5 51.5 49 cm
49.4 20.27 19.29 Inches
Table 3-15 Mass of Venue Go™ without Cart and Peripherals
Model Mass [KG] Mass [LBS]
Venue Go™ 6 13.22