D
IRECTION GB091046, REVISION 2 VIVID E9 / VIVID E7 BT’13 SERVICE MANUAL
Chapter 8 - Replacement procedures 8 - 11
8-4-8 When to load or reload the software
The software loading procedure is somewhat different, depending on why you need to load the software.
Use the table below to make the choice.
Table 8-4 Installation choices
Why Start here
• The software installed on the VIVID E9 / VIVID E7 has become corrupted.
• VIVID E9 / VIVID E7’s configuration/setup has issues that are difficult to
troubleshoot and resolve, so it may be easier to do a software reload, and
start the setup from scratch.
8-4-9 "Reloading the Software from Repository" on page 8-12.
The same as above, but it is impossible to access OLC.
8-4-10 "Loading the software from USB Flash Drive (UFD)" on
page 8-15
• You want to erase all data on the HDD
• BEP or HDD has been exchanged.
• The other methods failed.
8-4-10 "Loading the software from USB Flash Drive (UFD)" on
page 8-15