EasyManua.ls Logo

GE Vivid T8 - Page 130

GE Vivid T8
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Setup
3-50 Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
3-9-4-2 Verify the Network Connection to a Device
Follow the steps below to do a First Test (TCP-IP Ping) of the
network connection:
1. Highlight the device to be verified and select Properties.
Figure 3-35. DICOM Server Properties dialog
2. Select the "Check" button to Ping the server.
NOTE: By selecting the "Check" button, a ping is sent to the remote
server to see if it is accessible via the network. It is not a
DICOM Echo (DICOM ping), so it does not check AE title or
port number.
3. If the network connection to the server is OK, it will be
illustrated by the “Passsign, a white check mark on a green
background .
The “Fail“ sign indicates that the network connection is
failing.
Typical causes:
Network cable not connected
Configuration error(s)

Table of Contents