Description of the System
Image Info:
B (2D) Mode
User program
7.5-5.0
Pwr -3
Gn -12
C5 / M7
P6 / E4
MI 0.4
TIS 0.0
Name of user program
Receiver frequency [MHz]
Acoustic power [dB]
Gain [dB]
Dynamic curve [number], and Gray map [number]
Persistence [number] and Edge enhancement [number]
Mechanical Index
Thermal Index
Image Info:
M Mode
Gn 10
150/C1
EE 1
Rej 10
Gain [dB]
max. Dynamic range [dB] and selected Dynamic curve [number]
Edge enhancement [number]
Reject [number]
Image Info:
D (PW, CW) Mode
Gn 10
WMF mid
SV Angle 60°
Size 2.5 mm
Frq mid
PRF 1.2 kHz
Gain [dB]
Wall motion filter [Hz]
Angle correction
Sample volume size
Transmit frequency [Probe data]
Velocity Range [kHz, cm/s, m/s]
Image Info:
CFM-Mode, PD-Mode
User program
7.5-5.0
Pwr -3
Gn -12
C5 / M7*
P6 / E4
MI 0.4
TIS 0.0
Name of user program
Receiver frequency [MHz]
Acoustic power [dB]
Gain [dB]
Dynamic curve [number] and Gray map [number]
Persistence [number] and Edge enhancement [number]
Mechanical Index
Thermal Index
Gn 60
Bal 120
Qual mid
WMF mid
PRF 1.2 kHz
Displ. V
Gain [dB]
Balance [number]
Quality of CFM [table]
Wall motion filter [table]
Velocity Range [kHz, cm/s, m/s]
Display Mode [wedge]