Technical Data / Information
7.4 Scan Converter
Video memory size: 800 x 600 x 32 bit
Image memory: 4 MB
Gray scale values: 256 (8 bit)
Depth range: depends on used probe
Image lines: max. 1024
Scan angle: max. 360°
Aspect ratio: min. 0.25:1 to max. 8:1 (Write 5:1, Read factor 0.8 to 2.4),
incremental magnification, no loss of resolution.
M-Mode: M-Mode search line positionable at each scan line
M-Mode depth range: same as B-Image
M-Mode full screen scroll time: 300 / 225 / 150 / 100 Pixels/sec. (50 Hz) in relation to monitor width
3.5 / 5.0 / 7.5 / 10 cm/s
2D/M-Mode simultaneous: yes (multi-element probes)
7.5 Cine Loop Memory
Capacity: up to 256MB
typ. 3000 2D-images
Call-up of sequence: manual image by image
automated: 50 to 100% of real-time rate, liberal selection of start- and end
image
7.6 Display Modes
2D-scan: Single, Dual, Quad,
Trapezoid-Mode, Focus and Frequency Composite (FFC),
Compound Resolution Imaging (CRI), Harmonic Imaging,
β-View (Beta View)
2D-volume scan: Volume Contrast Imaging (VCI)
3D-scan: Multiplanar analysis, Volume scan
4D-scan: Real-time 4D, Real Time 4D Biopsy
M-Mode: 2D/M
Doppler-Mode: 2D/D horizontal Partition, three different formats 40/60, 50/50 and 60/40
Color Doppler Mode: 2D/CFM (Single, Dual, Quad), 2D/CFM/D, 2D+2D/C
Image orientation: left/right and up/down selectable