Technical Data / Information
Ambient temperature: 10°C to 40°C resp. 50°F to 104°F (operation temp. of instrument)
-10°C to 40°C resp. 14°F to 104°F (stock and transport temp.)
working temp. of probes see probe user’s manual
Barometric pressure: 700 to 1060 hPa (operation condition)
700 to 1060 hPa (stock and transport condition)
Humidity: 30 to. 80% RH no condensation (operation condition)
0 to. 90% RH no condensation (stock and transport condition)
Humidity protection: covered, no humidity protection
Over-temperature protection: yes, provided by three fans
Dimensions: 680 x 1000 x 1450 mm [width x depth x height]
26.7 x 39.4 x 56.1 in.
Weight: basic unit (without accessories) approx. 136 kg
7.2 Transmitter
Frequency range: Broad band system 1 to 16 MHz, automated adaptation to the probe
used.
Control of acoustic output: Range: 32 dB, adjustable in 1dB steps
Focusing: selectable transmit focal length and focal depth
Processing-channels: 512 channels
Sound field parameters: the declaration of the sound field parameters acc. to IEC1157 can be
furnished by the manufacturer on request. Sound field limiting
parameters on request.
7.3 Receiver
Frequency range: Broad band system 1 to 16 MHz, automated adaptation to the probe used.
Focusing (with annular and multi-element):
Sub-pixel based digital dynamic focusing system:
Accuracy of focus: +/- 3ns
Sample-rate: 60.0 MHz
Processing-channels: High resolution Mode: 512 channels
Receive-apodization: yes
TGC: manual, 100 dB control range by gain knob and slider-pots
Dynamic range: 150dB
Voluson
®
730 - Instruction Manual
7-4 105838 Rev. 3