DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
5-74 Section 5-17 - LINUX - Voluson Maintenance System
13.)Please wait until all processes are finished and the final “DOC” window (see figure below) appear.
14.)Press A
key to activate the Ultrasound Application Software [UIS].
15.)If prompted, enter the “Permanent key” and then click Submit.
16.)Customize the system as described in Section 5-17-8 on page 5-80.
Figure 5-49 press the “A” key to start UlSApp
Figure 5-50 enter Permanent key