EasyManua.ls Logo

Genevac HT-24 - Page 45

Genevac HT-24
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Manual
Genevac HT-24
10-1562 Issue 3-7 March 2014 43
Example 2. Fast Lyophilisation of HPLC Fractions
Method consists of three runs linked together to form a multi stage run, with settings as
follows:
Stage 1
Ramp & Remove
Organic Phase
Stage 2
Freeze & Primary
Drying
Stage 3
Secondary
Drying
Run No.
1
2
3
Run Data Protection
N
N
N
Next Run Number (0 for none)
2
3
0
Run Name
Fractions Lyo
Fractions Lyo 2
Fractions Lyo 3
Sample Holder Type
Your Holder
Your Holder
Your Holder
Sample Volume (ml)
Your Volume
Your Volume
Your Volume
Solvent
MeOH & H
2
0
MeOH & H
2
0
MeOH & H
2
0
SampleGuard Control Temp (°C)
40
0
40
SampleGuard Control Channel
1
1
1
CoolHeat Enable Pressure (mbar)
50
50
50
Chamber Temperature control (°c)
preheat to 20
preheat to 0
preheat to 30
Rotor Speed
High
High
Low
Heat-Off Option
Elapsed Time
Elapsed Time
Elapsed Time
Heat-Off Elapsed Time (HH:MM)
00:45
0:00
05:00
End-of-Run Option
Elapsed Time
Elapsed Time
Elapsed Time
End-of-Run Elapsed Time (MM:MM)
00:45
03:00
15:00
Auto Defrost & Drain
N
N
Y
Pressure Control Regime
Variable Dri-Pure
Full Vacuum
Full Vacuum
Dri-Pure Start Pressure (mbar)
120
-
-
Dri-Pure End Pressure (mbar)
1
-
-
Dri-Pure Ramp Time (MM)_
45
-
-
Final Control Pressure (mbar)
0
-
Note: There are a number of variables that can significantly affect the timings of
lyophilisation methods. For assistance in optimising fast lyophilisation methods, contact
applications@genevac.co.uk

Table of Contents

Related product manuals