10-1562 Issue 3-7 March 2014 53
Continued
See IR lamp / fan failure above
See SampleGuard error above
Check vacuum connectors. Contact
Genevac service if leak cannot be
found
Aqueous samples freezing
due to evaporative cooling
Set controlled pressure to 8 mbar for
aqueous runs
Defrost / drain condenser
Waste solvent container
valve closed
Open the waste solvent container valve
(see: Waste solvent container)
Possible failure of condenser drain
pump or blockage of drain pump filter.
Contact Genevac Service for
assistance.
Samples not dry on lower
rotor levels only
Go to OPTIONS menu and enable
lamp layers
Condensation inside
chamber
Chamber walls are too cold.
Refer to: Condensation
Set chamber temperature control to
pre-heat chamber (specify
temperature)
Go to Run Edit screen and select the
Dri-Pure option from the Pressure
Control Regime drop down menu
Set chamber temperature control to
wait to cool (specify temperature)
Pump purge before power-
down ā HCl systems only
No fault, this is normal. Wait 45
minutes for pump to purge before
powering down system