6.2.3.1.3 Refit the spray arm
003146
2
1
3
1
4
1. Bearing ring
2. Center nut, with lug
3. Center nut, without
lug
4. Spray manifold
After cleaning, refit the spray arm as follows:
1. Refit the plastic ends on the spray arm and press the metal
clip back into the groove on each side.
2. Refit the spray arm to the spray manifold.
The upper spray arm must be fixed with the lug of the
center nut facing upwards
Ensure that the bearing ring covers the spray manifold.
3. Check that the spray arm can rotate freely.
6.2.3.2 Clean the spray manifold
1. Remove the spray arm manifold from the chamber wall by
loosening the nuts.
2. Remove the spray arm from the spray arm manifold, see
section 6.2.3.1.2 Remove and clean the spray arm.
3. Using a bottle brush, clean the inside of the spray arm
manifold.
4. Refit the spray arm manifold.
5. Refit and check the spray tube, see section 6.2.3.1.3 Refit the
spray arm.