OPERATION -
27
The fault codes are explained later in this chapter.
Blowing safety valve
If a safety valve blows off:
Press the EMERGENCY STOP button.
If this has no effect, cut off the electric power supply with the control
switch on the front of the unit.
In an emergency, the power supply can be cut off with the working
switch.
Send for a service technician without delay.
An alarm signal goes off and the relevant fault code is
displayed on the cycle screen, while the cycle screen’s
background turns red.
After the alarm signal goes silent, the
Alarm indicator in the status field blinks.
If the unit is equipped with a printer, the alarm text is
also printed out as part of the cycle log.
In case of alarm:
Read the error message and correct the fault.
In case of an error concerning a temperature sensor, a
pressure transducer or the Back-up battery a service tech-
nician is to be called for.
On power failure or if the control switch or working switch
of the sterilizer had to be turned off during a process, there
is a risk of leakage at the door. Safety measure when there
is positive pressure in the chamber:
KEEP WELL AWAY FROM THE DOOR AND CALL A TECH-
NICIAN!