OPERATION -
29
Call a technician.
In case of a serious fault, always call a technician.
After the technician activates the key switch to the Stepping/Authorized
user setting, the technician has the following alternatives:
Restart the cycle where it stopped.
Move the cycle to another phase and start the cycle from there or
move it to the end.
Shutting off the alarm indicator - Avanti.
If the unit is equipped with a printer, the alarm text is
also printed out as part of the cycle log.
To restart:
Press
Start
.
To move to a different step:
Press
More
and then repeatedly press
Step
.
After the cycle ends:
The Finished Cycle indicator lights up red after the
cycle is finished.
Acknowledge the alarm by pressing
More
and then
Ack.
If the unit is equipped with a printer, the alarm text is
also printed out as part of the cycle log.
For safety reasons, it is not possible to step past a pres-
sure equalisation phase or a cooling phase.
Finish the current cycle.