3.3.1 Preparations and dismantling
Unpack the O
2
cell at least 15 minutes
before replacement. Turn the ventilator off
using the switch on the rear of the user
interface.
Disconnect the ventilator from the power
and gas supply.
SERVO-i
SERVO-i
Lift the locking handle and pull out the
patient unit.
SERVO-i
Lift the handle and fold it over the Expiratory
cassette.
Turn the bayonet screw one quarter of a
turn to release.
Lift off the cover.
28