4.2.3 Emergency opening of the machine
WARNING!
Goods in the machine MUST be reprocessed in
the appropriate process after an Emergency
Opening is performed.
WARNING!
If the goods are still hot after a program is
completed, handle them with care to avoid burns.
In case of power failure, you can open the machine using the cord
attached at the right, inside the lower front door.
Remove the screws.
Unhook the door and fold it downwards.
Pull the cord and the door opens.
1
002236
Figure 4. Cord for emergency opening
1. Door that is folded downwards. There is a emergency
opening cord behind the door.
4 ALARM MANAGEMENT
54 (128) TECHNICAL MANUAL GETINGE WD15 CLARO