A30 High temperature in
wash chamber
The temperature in the chamber is more than 5°C above the
highest reference point.
A31 External communication
error
DS30 Netcom is selected and communication with Netcom is
not
working. (If SF215 is set.)
Possible action:
a. Check the cabling.
b. Check the settings
A32 Log error DS30 Netcom is selected and the log is not working. (If SF217
and
SF218 are not set 2 min. after end of process.)
Possible action:
a. Check the USB memory stick
b. Check the settings
A33 Low disinfection
temperature
The temperature in the wash chamber is below the set value
for disinfection
(for a period longer than 60 seconds).
Possible action:
a. Check that there is power on all phases up to the
elements.
b. Check the calibration of the temperature sensors
4 ALARM MANAGEMENT