Instruction manual Installation
11
The U1000 V2 boots for the first time.
The GF start screen is displayed for 5 s.
The start-up wizard automatically starts.
Enter pipe id menu appears.
Select your pipe inside diameter and press
to acknowledge.
See Section 9.1 Entering the pipe inside
diameter, Page 19 for more details.
Select your pipe material and press to
acknowledge.
Select your instrument fluid and press
to acknowledge.
Select you media temperature:
- 'Cold' for temperature 2 °C to 40 °C
- 'Hot' for temperature 40 °C to 85 °C
Press to acknowledge.
The U1000 V2 shows you the correct
transducer distance for your application.
Press to acknowledge
Proceed with 8.4 Adjusting the sensor
distance
The U1000 V2 searches for a valid signal.
If a valid signal is detected, the signal strength
and flow are displayed.
The direction of flow at switch-on time is taken
as a positive flow value and the current and
pulse output refer to this flow direction.