Series 8035XA Peak Power Sensors
1-4 Manual 21568, Rev. F, March 2008
Zero Accuracy
Applies to 80350A, 80353A and 80354; 80351A = 100x larger,
80352A = 1000x larger
Zero Set: < ±1.0 mW, Peak;
< ±0.05 mW, CW
Zero Drift: < ±1.0 mW, Peak
< ±0.05 mW, CW
in 1 hour at constant temperature, 24 hour warmup
Noise Uncertainty: < ±1.0 mW, Peak;
< ±0.05 mW, CW at constant temperature, measured over a 1
minute interval, 24 hour warmup
Sample Delay Timing
Delay Range: -20 ns to 104 ms
Delay Resolution: 0.5 ns
Delay Jitter: ±2.0 ns
Trigger Level Set Range:
Internal: -30 to +20 dBm
Resolution: ±0.01 dB
External: 0.0V to 4.0V
Resolution to 0.01V
Trigger Jitter: < ±2.0 ns
Settling Time: (50% to within 3%) < 250 ns
Control Inputs and Outputs
Trigger Input: [SMB (jack) connector]
TTL External Trigger Input (absolute maximum = 10 V)
110 kΩ Impedance.
Detector Out: [SMB (jack) connector]
Monitor real time pulse waveform on an oscilloscope with this
voltage output (uncalibrated)
(High Impedance - Do Not Terminate)
Sample Delay: [SMB (jack) connector]
High, 5 V, between trigger and sample points. Connect to digital
oscilloscope channel 2 for triggering and sample point
identification
(High Impedance - Do Not Terminate)
Maximum cable length = 3 meters