12
3) BP extending tube
4) A pouch
5) User Manual
6) Software
Note
The monitor also can be equipped with pediatric and neonatal cuff, if necessary, please contact our
company or its representatives.
The width of the cuff should be 40% of the limb circumference(50% for the newborn) or 2/3 of
the length of the upper arm.The length of the inflated part of the cuff should be sufficient to
surround 50% to 80% of the limb. Unsuitable cuffs can produce erroneous readings. If there is
a problem with the size of the cuff, use a larger cuff to reduce the error.
Reusable Cuff of adult/ pediatric/ neonatal
Patient type
Limb circumference
Width of the cuff
Length of inflatable tube
Neonatal
6
11 cm
4.5 cm
1.5 m or 3 m
Infant
10
19 cm
8 cm
Pediatric
18
26 cm
10.6 cm
Adult 1
25
35 cm
14 cm