The room temperature rise or fall sharply, or the environment of measurement often changes;
Measuring in a moving vehicle;
The position of cuff applied (higher or lower than the heart level);
Continuous measurement for a long time;
4.4 NIBP Error Messages and Solutions
Display message
Cause
Solution
Low battery Device battery is low.
Replace the battery. If the problem
still exists, please contact us.
Loose cuff Cuff is not connected correctly.
Reconnect the cuff. If the problem
still exists, please contact us.
Atmospheric
pressure error
Valve can not be open.
Restart the device. If the problem still
exists, please contact us.
Signal is too
weak
Object measuring the pulse is too
weak or the cuff is loose.
Check the cuff connectio
n, tighten
the cuff if it is loose.
It is over the
range
Object measuring blood pressure is
over the measurement range.
Take another measurement.
If the
problem still exists, please contact us.
Excessive
movement
Movement may result in too much
interference in the signal during
Be sure to keep still during
measuring process.