measuring process.
Overpressure
Cuff pressure is over the scope,
ADU 300 mmHg, Newborn:
150mmHg.
Check the cuff to make sure it is not
blocked or squeezed.
Saturated signal
Movement or other factors may
lead to too big signal amplitude.
Check the connection of air tube to
make sure it is not squeezed. Patient
should keep quiet and then take a
new measurement.
Air leakage
Possible air leakage in the valve or
airway
Check the air tube and the cuff.
System failure
Possible failure caused by pump,
air valve or pressure sensor.
Please contact us.
It spends too
much time
The time for a single measurement
exceeds the maximum
measuremen
t time (adult: 180s,
neonate: 90s).
Check the connection of air tube and
tighten the cuff.