Options for "AWAKE INTERVAL(min)" and "ASLEEP INTERVAL(min)": 5,10,15, 20, 30, 40, 60,
90, 120, 180, 240;
The step of each adjustment for "AWAKE TIME" and "ASLEEP TIME" is 30 minutes. The
adjustment range: 00:00~23:30.
Note
The set measurement interval in "AWAKE INTERVAL" and "ASLEEP INTERVAL" is time
interval when automatically start the measurement under the mode of ambulatory blood
pressure, not including manual starting. For example: set "AWAKE TIME" to 7:00, set
"AWAKE INTERVAL" to 15min, then, the device will make the first blood pressure
measurement at 7:15; if the user start a blood pressure measurement by pressing measuring
button between 7:00-7:15, the device will also automatically start the measurement at 7:15,
and not affected by manual measurement.
After each item of this interface is set, the ambulatory blood pressure menu also need to be correctly
set to start the ABPM function. Refer to 3.4.5 ambulatory blood pressure menu for details.