The pressure by cuff may cause temporary weakness of some functions of the body. So do not
use monitoring medical electrical equipment on corresponding arm.
Do not move during measuring, because it will have a delayed effect on the patient's blood flow.
The device needs 2 hours recovery time to reach its performance of intended use after taking
out from the lowest storage temperature.
The device needs 4 hours recovery time to reach its performance of intended use after taking
out from the highest storage temperature.
1. Plug the air hose to the cuff socket on the device, and connect the device to power supply.
2. Apply the cuff to the patient's upper arm following the instructions below (Figure4.2.1).
Ensure the cuff is completely deflated.
Apply the appropriate size cuff to the patient, and make sure that the symbol "φ" is over the
appropriate artery. Ensure that the cuff is not wrapped too tightly around the limb. Excessive
tightness may cause discoloration and eventual ischemia of the extremities.
Figure 4.2.1 Cuff applying