Guidance and manufacturer's declaration – electromagnetic immunity –
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of device should assure that it is used in such an environment.
Immunity test
IEC60601 test
Compliance
Electromagnetic
Electrostatic
discharge (ESD)
IEC61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
Power frequency
(50/60Hz)
magnetic field
3A/m
3A/m
Mains power quality should
be that of a typical
commercial or hospital
NOTE: U