Guidance and manufacturer's declaration electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer's declaration electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of device should assure that it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
3A/m
3A/m
Mains power quality should
be that of a typical
commercial or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.