EasyManuals Logo

Global Surgical SMR MAXI Standard User Manual

Default Icon
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
Section 1 Important Safety Instructions
1-3
WARNING
!
THIS SYSTEM IS INTENDED FOR USE BY HEALTHCARE PROFESSIONALS
ONLY.
WARNING
!
THIS SYSTEM MAY CAUSE RADIO INTERFERENCE OR MAY DISRUPT THE
OPERATION OF NEARBY EQUIPMENT. IT MAY BE NECESSARY TO TAKE
MITIGATION MEASURES, SUCH AS RE-ORIENTING OR RELOCATING SMR
CABINET OR SHIELDING THE LOCATION.
WARNING
!
TOO HIGH OF PRESSURE COULD RESULT IN HAZARDOUS OUTPUT FOR
DISPENSING MEDICATIONS.
CAUTION
!
A Hydrophobic Filter is installed in the vacuum line. Failure to replace this lter
regularly (at least once a month) will lead to reduced ow. Extended use of
alcohol or solvents may reduce the e󰀩ciency of the microporous membrane of
the hydrophobic lter. If these chemicals are used frequently, the lter should be
inspected and replaced more often than the recommended time.
CAUTION
!
Non-use of the hydrophobic lter may lead to possible contamination and pump
failure. Global Surgical will not receive or replace any pump in which the
Hydrophobic Filter was not used.
CAUTION
22
Canister is Single Use Only. Do Not attempt to clean, sterilize or reuse canister.
Possible consequences of reuse include: 1.) implosion, 2) uid bypass, and
3) exposure to blood borne pathogens.
CAUTION
!
Canister is Single Use Only, Do Not attempt to clean, sterilize or reuse canister.
Possible consequences of reuse include: 1.) Implosion, 2) Fluid bypass, and
3) Exposure to blood borne pathogens.
CAUTION
!
Care should be used by all personnel coming into contact with suction
equipment. Wear gloves when handling suction equipment. Hand washing is
necessary to prevent the possibility of cross-contamination.
CAUTION
Dispose of the disposable suction canister containing medical waste in an
environmentally safe manner per state and federal regulations.
CAUTION
!
After moving cabinet to its permanent location, lock the four casters to keep it
from moving.
CAUTION
!
DO NOT block the Light Source ventilation slot on the back side of the MAXI
Treatment Cabinet during operation.
CAUTION
!
Drawers must be closed before transporting any MAXI Treatment Cabinets.
CAUTION
!
The internal power receptacles located inside the cabinet are not for general
use.
CAUTION
!
Failure to cut and remove the shipping tape from around the two pumps will
result in damage to the pumps.
CAUTION
To reduce the risk of electric shock do not open pump compartment while mas-
ter power switch is turned on. Refer servicing to qualied service personnel
CAUTION
!
To prevent tipping, push the cabinet from the side not the front or back. Only
open one drawer at a time.
CAUTION
!
Before moving cabinet to a di󰀨erent location, remove all loose equipment
from the counter top.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Global Surgical SMR MAXI Standard and is the answer not in the manual?

Global Surgical SMR MAXI Standard Specifications

General IconGeneral
BrandGlobal Surgical
ModelSMR MAXI Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals