NOTE
The control solution range printed on the test strip vial is for control
solution use only. It is not a recommended range for your blood test
levels.
For any control solution test, your device will tag this measurement
as QC test automatically.
● To avoid contaminating control solution, do NOT directly apply
control solution onto a test strip.
Important Control Solution Information
● ONLY use TAIDOC’s control solutions with your meter.
Do not use the control solution beyond the expiration date or 3
months after first opening. Write the opening date on the control
solution vial and discard the remaining solution after 3 months.
● It is recommended that the control solution test to be done at room
temperature between 20°C to 25°C (68°F to 77°F). Please make
sure your control solution, meter and test strips are at the specified
temperature range before testing.
● Shake the vial before use, discard the first drop of control solution
and wipe it off the tip to ensure a pure sample and an accurate result.
● Store the control solution with the cap tightly closed at temperatures
between 2°C to 30°C (35.6°F to 86°F). Do NOT freeze.