13-4
OPERATOR MAINTENANCEUSER HANDBOOK
4) Inspect the hydrophobic lter for contamination or damage.
If either is evident, replace the lter.
5) Fit a new Hydrophobic Filter Bulb (Part No. 67213), if
required.
Note: Installation is a reversal of the removal procedure
and therefore relevant Figs. should be referred to.
6) Connect the hydrophobic lter bulb to the probe lter bulb
so that the bayonet connection locates correctly then turn
hydrophobic lter bulb in a clockwise direction until secure.
Note: Care must be taken not to overtighten the filter
assembly.
7) Before attaching the probe adaptor to the hydrophobic
lter bulb, check the probe washer for signs of damage.
If damaged, it must be replaced with a new washer
(Part No. 12379).
8) Hold the hydrophobic lter bulb then attach probe adaptor to
the hydrophobic lter bulb by turning adaptor in a clockwise
direction until secure.
Note: Care must be taken not to overtighten the probe
adaptor.
9) Attach the lter holder complete with new dust lter as
detailed in section 13.2.1 paragraphs 3 to 5.
10) Switch the instrument ON and check that sample / ow fault
is displayed when the probe inlet is blocked (with a nger for
example) while the pump is running. If fault is not displayed,
check tightness of all ttings.