4-1
LEAK TEST MODE
This mode is for the technician to investigate odour or leak complaints
and to pinpoint the leak.
Note: The instrument pump must be switched ON to measure
in Leak Test mode.
4.1 LEAK TEST RANGES
Leak Test mode will have the following ranges available for normal
operation:
0 – 2000 ppm Methane (CH
4
) Flammable with following
options:
(to display as LEL with 0.1% resolution up to 9.9%).
(to display without numerical indication).
(to display zero until gas reading exceeds 100 ppm).
0 – 100% LEL Methane (CH
4
) Flammable with option:
(to display as gas in air).
0 – 100% VOL Methane (CH
4
) Flammable.
0 – 2000 ppm Carbon Monoxide (CO) if tted.
4.2 LEAK TEST FEATURES
Leak Test mode will have the following features available:
Audible / Visual PPM Ticker (Geiger) Indication
Audible / Visual Gas Alarms
• Backlight
• Flashlight
Pump - can be toggled ON / OFF and will stop if ow fault
is detected
Max Display
Autoranging from ppm-LEL-Gas
Manual Range selection between Flam and Flam plus CO
(Dual Display)
Threshold adjustment
Display Invert
Automatic datalogging