EasyManua.ls Logo

GMI GT series - Confined Space Mode Features; CSM Features

GMI GT series
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-1
CONFINED SPACE MODE
This mode is used for conned space pre-entry testing and for
personal monitoring in areas such as basements etc.
5.1 CSM RANGES
Conned Space Mode (CSM) will have the following ranges available:
0 – 100% LEL Methane (CH
4
) Flammable
(EEE Over-range) and option to display up to 9.9% LEL with
0.1% resolution
0 – 25% Oxygen (O
2
) if tted
0 – 2000 ppm Carbon Monoxide (CO) if tted
0 – 100 ppm Hydrogen Sulphide (H
2
S) if tted
5.2 CSM FEATURES
Conned Space mode will have the following features available:
Audible / Visual Alarms.
Min / Max Display.
• Backlight.
• Flashlight.
Pump ON continuously - will not stop if ow fault is detected
- ow fault alarm will be activated.
Datalogging – automatic every 60 seconds.
Manual Logging is also available. This will log all gas
concentrations at that instant.
Time weighted averaging for Alarms.
Condence signal every 15 seconds.
Display invert.
5