EasyManuals Logo

GMI GT series User Manual

GMI GT series
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #111 background imageLoading...
Page #111 background image
13-9
OPERATOR MAINTENANCEUSER HANDBOOK
9) Before attaching the lter adaptor to the lter housing,
check the O-ring shown in Fig. 13.4 for signs of damage. If
damaged, it must be replaced with a new O-ring (Part No.
12737).
10) Attach lter adaptor to lter housing, in a clockwise direction,
then tighten to secure. See Fig. 13.4.
Note: Care must be taken not to overtighten the adaptor.
11) Before attaching the lter housing to the lter bulb, check the
washer for signs of damage. If damaged, it must be replaced
with a new washer (Part No.12379).
12) Hold the hydrophobic lter bulb then attach chemical lter
to hydrophobic lter bulb by turning in a clockwise direction
until secure. See Fig. 13.3.
Note: Care must be taken not to overtighten the chemical
lter.
13) Attach probe adaptor to chemical filter, in a clockwise
direction, then tighten to secure. See Fig. 13.3.
Note: Care must be taken not to overtighten the probe
adaptor.
14) Attach the lter holder complete with new dust lter then
perform leak check, as detailed in section 13.2.1 paragraphs
3 to 6.

Table of Contents

Other manuals for GMI GT series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GMI GT series and is the answer not in the manual?

GMI GT series Specifications

General IconGeneral
BrandGMI
ModelGT series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals