EasyManuals Logo

GMI GT series User Manual

GMI GT series
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
9-4
SNIFFER MODEUSER HANDBOOK
9.5.3 Zero
To zero the ppm range
Press and hold the RH button .
Note: The pump must be switched ON to zero the ppm range.
9.5.4 Flow Fault
If ow fault is detected, the pump stops automatically. The instrument
should be checked for water ingress or blockage.
To clear the ow fault:
(Once the blockage has been cleared):
Press the RH button .
9.5.5 Ticker (Geiger) On / Off
To enable / disable the Geiger (Ticker):
Press and hold the UP and DOWN buttons
simultaneously, as follows:
1. Press and hold - to enable visual only.
2. Press and hold again - to disable both audible and visual.
3. Press and hold again - to enable both audible and visual.
9.5.6 Ticker (Geiger) Adjust
To adjust the Ticker (Geiger) threshold:
(after the Ticker (Geiger) feature has been enabled)
Press the UP or DOWN buttons.
The threshold setting is displayed briey beneath the PPM CH
4
.

Table of Contents

Other manuals for GMI GT series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GMI GT series and is the answer not in the manual?

GMI GT series Specifications

General IconGeneral
BrandGMI
ModelGT series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals