ALARMSUSER HANDBOOK
The following Table 12.2 shows the alarm options available in the
Leak Test and Conned Space Modes.
LEAK TEST CSM
FLAM CO H2S O2 FLAM CO H2S O2
WARNING
20
Hi
Hi Hi 20 35 15 23
STEL 200 10
LTEL 30 5
Lo
Lo Lo
19.5
Table. 12.2 Alarm Options
Note:
1. Where there is a square with a number, it means that the GMI
default is to have these alarms active at the gas concentration
shown by the number.
2. Where there is a square without a number, it means that
this alarm option is available should the user wish to have it
activated.
3. The alarm concentrations must, if active, be the same level
in both modes.