EasyManuals Logo

GMI GT series User Manual

GMI GT series
147 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
12-10
ALARMSUSER HANDBOOK
The faulty sensor will cause
the instrument to display a
flashing spanner symbol to
warn the user that this sensor
is not working correctly, as
shown in Fig. 12.8:
12.6.3 Sensor Fault
There are two types of sensor fault as illustrated in the following
displays:
1) If a “ZERO FAULT”
flag and a flashing
spanner symbol
appear, alternating
with a zero reading
as shown in Fig. 12.9,
apply relevant test
gas (LEL in example)
for two minutes and
allow the display to
return to zero, then
switch instrument Off
and On again.
If fault remains,
return instrument to
an approved Service
/ Repair facility.
alternating with
Fig. 12.8 Zero Fault
alternating with
Fig. 12.9 Sensor Fault

Table of Contents

Other manuals for GMI GT series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GMI GT series and is the answer not in the manual?

GMI GT series Specifications

General IconGeneral
BrandGMI
ModelGT series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals