-59-
minutes, monitoring will resume in the same session when the monitor is turned back on. If the moni-
tor is turned off for more than fteen minutes, it will begin the next session when it is turned on again.
Tension Guides. Extensions which connect the penile loops to the RigiScan Monitor.
Tumescence. The circumference of the penis, measured in centimeters.
Tumescence Activity Units (TAU). This value is a time-intensity measurement that represents the
area under the tumescence curve above the baseline during qualied events. It is proportional to the
percent increase in tumescence over baseline. The Tumescence Activity Unit is calculated by summing
the tumescence value minus the baseline tumescence and dividing by 4 times the baseline. The value
4 is used in the calculation because there are four tumescence samples taken per minute.