Guidance and manufacturers declaration - electromagnetic emissions
The RigiScan is intended for use in the electromagnetic environment specied below. The customer or the user of
the RigiScan should assure that it is used in such an environment.
Emissions
Test
RF emissions
CISPR 11
The RigiScan uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The RigiScan is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Group 1
Group B
TBD
TBD
RF emissions
CISPR 11
Harmonic
IEC 61000-3-2
Voltage uctuations/
icker emissions
IEC 61000-3-3
Compliance Electromagnetic environment — guidance
Guidance and manufacturers declaration - electromagnetic immunity
The RigiScan is intended for use in the electromagnetic environment specied below. The customer or the user of
the RigiScan should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61.000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Power frequency
(50/60 Hz) magnetic
eld IEC 61000-4-8
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
± 6 kV contact
± 8 kV air
3 A/m 3 A/m
±2 kV for power
Supply lines
± 1 kV for
input/output lines
± 1 kV line(s) to
line(S)
± 2 kV line(s) to earth
<5 % (IT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in U-1)
for 5 cycles
70 % (IT
(30 % dip in UT)
for 25 cycles
<5 % UT 180°
(>95 % dip in UT)
for 0.5s
<5 % (IT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in U-1)
for 5 cycles
70 % (IT
(30 % dip in UT)
for 25 cycles
<5 % UT 180°
(>95 % dip in UT)
for 0.5s
± 1 kV line(s) to
line(S)
± 2 kV line(s) to
earth
±2 kV for power
Supply lines
± 1 kV for
input/output lines
Compliance
Electromagnetic environment
– guidance –
Floors should be wood, concrete, or ceramic tile. If
oors are covered with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
RigiScan requires continued operation during power
mains interruptions, it is recommended that the
RigiScan be powered from an uninterruptible power
supply or a battery.
Power frequency magnetic elds should be at levels
characteristic of a typical location in a typical com-
mercial or hospital environment.