CE6000-UM-251-9370 7-17
7 ELECTRICAL ADJUSTMENTS
7.9 Adjusting the Registration Mark Sensor Sensitivity
This adjustment will set the registration mark sensor sensitivity.
The sensitivity of sensor need adjust if you replace the main board or the registration mark sensor.
How to adjust the registration mark sensor sensitivity
(1) Turn on the power while pressing the FAST and ENTER keys.
(2) Load an A4 size sheet of paper into the plotter.
The paper color must be white, and it recommend copy paper.
(3) Select the Language and the Length Unit if the Language selection menu is displayed (See section
7.5.).
(4) Press the PAUSE/MENU key at the following menu.
(5) The following menu is displayed, and then press the right arrow key (ADJ).
(6) The following menu is displayed.
(7) Press the up arrow key until the following menu is displayed.
(8) Press the F1 key (RM SENSOR LEVEL ADJ). The following menu is displayed.