EasyManua.ls Logo

GRAPHTEC GL450 - External Input;Output Functions

GRAPHTEC GL450
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
5-7
External Input/Output Functions
Item Description
Input/output types Trigger input (1 ch), Logic input (4 ch), Pulse input (1 ch),
Alarm output (4 ch)
Input specifications Maximum input voltage: +24V
Input threshold voltage: Approx. +2.5V
Hysteresis: Approx. 0.5 V (+2.5 to +3 V)
Alarm output specifications Output format: Open collector output (100 k pull-up resistance)
Output conditions: Level judgment, window judgment,
logic pattern judgment, pulse judgment
Pulse input Revolutions mode (engines, etc.)
• Function: Counts the number of pulses per second; enables them to be
converted to rpms.
• Ranges: 500, 5 k, 50 k, 500 k Revolutions/F.S.
Counts mode (electric meters, etc.)
• Function: Displays a count of the number of pulses for each sampling
interval from the start of measurement.
• Ranges: 50 k, 500 k, 5 M, 50 M, 500 M C/F.S.
Inst. mode
• Function: Counts the number of pulses for each sampling interval.
Resets the count value after each sampling interval.
• Ranges: 50 k, 500 k, 5 M, 50 M C/F.S.
Maximum number of pulse inputs
• Counts, Inst. modes: 50 k/sampling interval
• Revolutions: 50 k/s